Skip to content
    Clinical Trials

    Open-face mask SGRT workflow maintained accuracy in head and neck cancer trial

    Ryan et alMethodsX

    Open-face mask SGRT workflow maintained accuracy in head and neck cancer trial
    Conventional thermoplastic immobilisation masks utilized in head and neck radiotherapy frequently cause patient anxiety and discomfort. Surface-guided radiotherapy (SGRT) offers non-ionising, real-time guidance for positioning and intrafraction monitoring, but clinical implementation must account for the independently mobile anatomy across the face, neck, and shoulders.
    This protocol describes the technical and clinical workflow developed for the SGRT-supported open-face immobilisation arms of the OPEN Phase III randomised controlled trial. The procedure begins with initial alignment using SGRT postural video guidance based on the patient's external contour. A facial setup region of interest (ROI) is used for head positioning. Subsequently, a larger monitoring ROI encompassing the face, neck, and upper thorax is applied to refine lower neck and shoulder positioning while maintaining facial alignment, as well as to perform continuous intrafraction monitoring throughout radiation delivery.
    The OPEN trial demonstrated that this workflow maintained setup accuracy, intrafraction stability, and clinical efficiency while supporting open-face immobilisation, providing a reproducible framework for clinical adoption.

    Key Takeaways

    • The OPEN trial demonstrated that the surface-guided radiotherapy workflow maintained setup accuracy, intrafraction stability, and clinical efficiency while supporting open-face immobilisation.